Onesummary:
Having a wide range of testing, covering measurements from conductor materials to insulation material testing;PC software for data processing and analysis, direct reading of resistance, resistivity, and conductivity data, with shielding protection function, integrated structural design, equipped with low resistance and high resistance measurement fixtures; 232 communication port, standard resistance calibration instrument; Provided in
Text or English operating interface
Conductor and insulation material resistivity tester
twoApplicable:
1. Suitable for measuring conductors and insulation materials
2. Suitable for wide range analysis of new materials
3. Used to analyze the process of material fusion from insulation to conductivity
4. Analyze the fusion of materials from conductors to insulation properties
5. Research institutes and R&D departments conduct research and development analysis
Conductor and insulation material resistivity tester
Parameter information:
Specification and model
|
FT-3204A
|
1. Resistance range
|
2*10^-8Ω(0.01uΩ)—2*10^16Ω
|
2. Range of resistivity
|
2*10^-8Ω.m—2*10^16Ω.m
|
3. Conductivity range
|
5*10^17~5*10^9s/m
|
4. Test current range
|
1. Low resistance range:100uA.1mA.10mA.100mA.1A.10A.;
2. High resistance range: 20uA 2uA200nA20nA2nA200pA20pA
|
5. Current accuracy
|
Error:100uA.1mA.10mA.100mA:0.1%;1A,10A:0.15%;
|
6. Voltage Range
|
1. Low resistance range:2mV.20mV.200mV.2V
2. High resistance range: 0-1000V
|
7. Voltage accuracy
|
2mV.20mV.200mV.2Vresolving power:0.1uV,1uV,10uV,100uV;
Range error:2mVfiles: 0.1%reading+ 6One word;20mV.200mV.2VFile:0.1%reading+ 2Character
|
8. Resistance accuracy
|
1. Low resistance: 0.01 μ Ω;≤0.3%
2. High resistance:<1G ±1% ;≥ 1G ±3%; ≥ 1T ±10% ;≥ 10T ±20% (Standard resistance)
|
9. PC software interface
|
PC software interface display: resistance, resistivity, conductivity, temperature, unit conversion, temperature coefficient, current, voltage, thickness, process map, report generation
|
10. Test mode
|
The range below ohms adopts constant current output mode; Ohms and above adopt constant voltage output mode
|
11. Working power supply
|
Input: AC 220V±10%.50Hz power consumption:<100W
|
12. Test fixture
|
Standard configuration includes one set of low resistance and one set of high resistance measurement fixtures
|
10. Calibration method
|
Two standard resistors (one low resistance and one high resistance)
|
Supporting plan: solve the status of various materials--Solid state, liquid state, gas state, granular resistance, resistivity, conductivity measurement
ROOKORuike Instrument-------Focusing on new material measurement and analysis instrument solutions
----Solution materials:
Electrical Performance Analysis of Conductors, Semiconductors, and Insulation Materials at Room Temperature and High Temperature Environments.
----We have been doing:
Research and development, production, sales. Research and development, production, sales, leasing, laboratory sample analysis, and extension services.