The ZN-YZJ-T pressure instrument automatic calibration system adopts original imported core components and applies multiple new technologies and processes, making the system very stable and reliable, creating a new concept of modern laboratory environment and giving it a brand new appearance. The product is accompanied by a verification certificate from the provincial metrology and testing department.
Performance characteristics
● Wide range: Built in detachable pressure module covers the full range, facilitating periodic calibration.
● Easy to operate: After setting through the panel keyboard, the required pressure value is automatically generated without the need for complicated processes such as operating a PC.
● Friendly interface: Large screen LCD display, Chinese menu, self-learning operation prompts, no need to refer to the manual.
Clean workbench: using gas and clean water as pressure transmission media and capable of automatically recovering accumulated water in the inspected table, achieving clean oil-free calibration.
Complete functionality: automatic storage of verification parameters, PC downloads verification data from the host as needed, and prints test reports.
Good stability: The pressure generation system is imported with original packaging, with no micro leaks, good stability, and high reliability.
Typical Uses
● It can calibrate primary and secondary instruments such as pressure gauges, pressure transmitters, pointer pressure gauges, differential pressure transmitters, pressure switches, etc. One calibration system is equivalent to a modern standard pressure laboratory.
Model Selection Table
A desktop structure designed according to ergonomics, with a metal frame and modular disassembly for easy transportation.
● Dimensions 1480 × 800 × 1350 (mm)
TECHNICAL INDEX
ZN-YZJ-T series high-precision digital pressure module | measuring range | Accuracy (level) | Power output 24V, 50mA Current measurement 0~21.000mA |
-10 ~0~ 10Kpa | 0.05 | ||
-95 ~0~ 100Kpa | |||
0~ 600Kpa | |||
0~ 6Mpa | |||
0~ 60Mpa |
Display and measurement technical indicators
Power output: DC24V 50mA
Current measurement: 0~21.000mA 0.05% F.S
LCD display: 240 × 128 dot matrix
Other options
● Computer hardware configuration:
1. PC: Lenovo P4-2.0G or above, 256M, 40G, optical drive, floppy drive, 17 "color display.
2. Notebook: IBM, P4-1.7GB or above, 256MB, 20GB, optical drive, USB flash drive, 14.1 "TTF.
3. Printer: HP color inkjet or Samsung laser printer.
Software configuration: Inspection report printing driver program.
Electromagnetic flowmeter K-type thermocouple thermal resistance bimetallic thermometer pressure transmitter flow transmitterThermal resistance verification simulator