Combining multiple functions - only 38cm x 50cm x 34cm in size, Appliskan is suitable for the most crowded laboratories with built-in oscillation and
Product details
  • Product Name: Appliskan Multi functional Reading Instrument
  • Order number: tlyon000388
  • Brand Name: Thermo, USA
  • Type: Multi functional ELISA reader
  • Filter position: None
  • Incubator: available
  • Printer: External
If you find that the product information is inaccurate, please feel free to correct it; For inquiries about related categories and brand products, please contact us

Thermo Fisher Scientific's Appliskan multifunctional microplate reader based on filters provides the best flexibility and will be the most popular new member for any life science laboratory. With this compact and small instrument, researchers can flexibly switch between multiple detection techniques without compromising detection performance. Combined with Thermo's powerful Skanlt software, experimental design, analysis, and reporting are carried out with its simple operation and flexible control. The built-in oscillation and incubation system, as well as the optional dual splitter and easy integration with automated systems, greatly enhance the high productivity of the laboratory.
Product Features
● Complete flexibility, whether it's now or in the future, whether it's your current or future needs, multiple measurement functions mean you only need a unique reading instrument - Aprisjan.
The compact instrument combines multiple functions - only 38cm x 50cm x 34cm in size, Appliskan is suitable for the most crowded laboratories.
Excellent sensitivity - you can be confident in your measurement results.
● Great plate compatibility - from large volume 6-well cell culture plates to small volume 384 well high-throughput plates.
Easy to use Skanlt software - accelerates experimental design, setting up and reporting even the most complex experimental measurements through an intuitive step list interface.
Accurate liquid separation - Apliskan can be equipped with up to two liquid separators to complete flash experiments and reduce the use of reagents.
Independent detection systems for different measurement techniques
The excellent performance of Appliskan in any application is based on independent detection systems using different measurement techniques. The high-sensitivity chemiluminescence mode uses specialized photomultiplier tubes to ensure excellent sensitivity and a wide dynamic range in any chemiluminescence application. Efficient xenon lamps and photomultiplier tubes with a wide wavelength range for fluorescence technology cover the wavelength range up to 820nm, providing high sensitivity and a broad linear dynamic range in all fluorescence measurement applications. The optical system specially designed for light absorption measurement covers wavelengths of 200-1000nm, ensuring extremely excellent linearity, accuracy, and repeatability in any colorimetric measurement, including DNA quantification.
Flexible and convenient choices
The flexibility possessed by Appliskan makes it easier for researchers to conduct research. Incubation, oscillation, built-in liquid separator, and simple filter operation mean the needs of each operator. For higher throughput measurements, Appliskan can be integrated into automation systems, such as the RapidStak microplate laminating machine for thermoelectric applications. Up to two optional liquid separators are convenient for adding reagents, and their effective separation and reagent emptying reflux functions can maximize the savings in experimental operating costs.
Powerful software with powerful functions
The Skanlt software for Appliskan features an intuitive user interface and a unique step list function. Skanlt software provides optimal instrument control, simple experimental design, and flexible data analysis and report formats to fully realize any of your applications.
Technical Parameter
light source
xenon lamp
detector
Fluorescence and chemiluminescence: PMT, light absorption: PDT
Detection type
Fluorescence intensity, chemiluminescence flash and glow, light absorption, time-resolved fluorescence (TRF), fluorescence polarization
Plate type
6-384 well plate
fluorescence
wavelength
200nm-1000nm
emission wavelength
360nm-820nm
Theoretical sensitivity
Fluorescence:<2 fmol Fluorescence/well, 384 well plate; TRF; <20amol Europium/well, 384 well plate
accuracy
Fluorescence polarization:<10mP1nM fluorescein, 96 well plate
dynamic range
>5 orders of magnitude (>5 decks)
detection time
Fluorescence: 5-1000ms, TRF and fluorescence polarization: 5-10000ms
light absorption
wavelength range
200-1000nm
linear
0-2.5Abs (96 well plate) at 450nm, ± 2%; 0-2Abs (384 well plate) at 450nm, ± 2%
measuring range
0-4Abs
resolution
0.001Abs
Time
20-1000ms
chemiluminescence
spectral range
Standard mode: 360-820nm; High sensitivity mode: 300-630nm
sensitivity
High sensitivity mode:<10 amol ATP/well, 384 well plate; Standard mode:<100mol ATP/hole, 384 hole plate
detection time
10-10000ms
linear range
>5 orders of magnitude
incubator
temperature range
Room temperature+4 ℃ to45℃
oscillator
Oscillation
Linear oscillation, 1-10mm
Liquid separator
quantity
Up to two
Maximum liquid separation volume
500ul (1000ul optional)
accuracy
<0.2ul or 2%, 5-500ul
accuracy
5-20ul<5%
Ordering Information
5230000
Appliskan,100-240V,50/60Hz
5230010
Appliskan with a liquid separator, 100-240V, 50/60Hz
5230020