Yancheng Dewang Instrument Equipment Co., Ltd
Home>Products>3140-A ultra-low emission flue gas analyzer
3140-A ultra-low emission flue gas analyzer
Analysis of high CO, high humidity, low SO2, and ammonia escape ultra-low emission flue gas components;
Product details

图片2.png



1. Scope of adaptation

ØtallCO、 High humidity, low SO2Ammonia escapeUltra low emissionsAnalysis of smoke composition;

ØMeasurement of efficiency of various desulfurization and denitrification equipment;

ØDetermination of oxygen content and excess air coefficient in flue gas(Optional)

ØVarious boilers and industrial furnacesSO2NO、NO2H2S、NH3Measurement of emission concentration and converted concentration;

ØOther applicable scenarios.

2. Main features

ØThe detection chamber adopts a Dawei prism folding optical path design, with long optical path, high gas resolution, and high detection limit;

ØLow temperature heating and high temperature detection modes are optionalapproveRemoving waterSO2NO2NH3absorbAdsorptionInterference makes the data more authentic;

ØBuilt in sampling pump, dust removal and zero gas generator, automatic zero calibration;

ØReal time detection of chamber temperature and pressure, and correction of detection results;

ØDesign of embedded micro thermal printer to achieve on-site printing of data;

ØEmbedded single board computer,Win7 operating system, dynamically displaying gas absorption curves;

Ø120GB solid state drive and large database communication software can archive monitoring data from previous years for future reference;

ØReal time display of the average value of monitoring data, doubleUSB interface, supports exporting data from USB flash drive;

ØExpand other UV gas monitoring projects without adding hardware;

ØAnti static design to avoid on-site static interference.

3. Main technical indicators

Project

Parameter range

resolution

Error

Smoke sampling flow rate(Suggestion)

1.0L/min

0.1 L/min

≤±2.5 %

AUTO ZEROsample flow

1.0L/min

0.1 L/min

≤±2.5 %

Zero adjustmentFlow control stability

≤± 2% (voltage varies between 180V~250V, resistance varies within 3-6 KPa)

Heating control temperature

40 ℃~160 ℃ (determined by actual needs)

1 ℃

≤±2 ℃

O2 (Electrochemical) Optional

0~25.0 %

0.1 %

Indication error: ≤± 2.5%

Repeatability: ≤ 2%

Response time: ≤ 30 seconds

Stability:24Within hours

Indication change: ≤ 1%FS

# SO2 (UV-DOAS method)

0200/1000mg/m3

1 mg/m3

#NO (UV-DOAS method)

0200/1000mg/m3

1 mg/m3

# CS2(UV-DOAS method)

050/500mg/m3

0.1 mg/m3

#Benzene (UV-DOAS method)

0~1000 mg/m3

1 mg/m3

# NH3(UV-DOAS method)

0~50.0/500mg/m3

0.1 mg/m3

#H2S (UV-DOAS method)

0200/1000mg/m3

1 mg/m3

electrochemistrySensor lifespan

Two in the airyear

Light source lifespan

Deuterium lamp 2000H, xenon lamp flashes 10 ^ 9 times

External dimensions

4U standard chassis

Instrument noise

<60 dB(A)

Whole machine weight

make an appointment12kg

maximumpower consumption

300 W(Turn on high temperature heating)

24-hour zero drift

low range1% FS; high range0.1%FS

nNote:#Indicating optional range items


Online inquiry
  • Contacts
  • Company
  • Telephone
  • Email
  • WeChat
  • Verification Code
  • Message Content

Successful operation!

Successful operation!

Successful operation!